Description
Applications
- Communications interface for programming and modification of operational parameters.
- The easily readable 4510 display can be used to monitor the process signal, simulate the output signal, and indicate sensor errors and internal device errors.
- The 4510 can be moved from one device to another. The individual system 3000 / 4000 / 9000 device configuration of a transmitter can be saved and downloaded to subsequent transmitters.
Technical characteristics
- Easy-to-read dot matrix LCD display.
- Backup memory for loading and saving of device configuration.
- Programming access can be blocked by assigning a password. The password is saved in the device in order to ensure a high degree of protection against unauthorized modifications to the configuration.
Mounting / installation / programming
- Mounting in Zone 2 / Div 2.
- All configuration data from a PR 3000 / 4000 / 9000 device can be transferred to a PC using the PR 4590.
- When mounted on devices that are installed upside down, a menu item allows the display on the 4510 to be rotated 180° and the up/down buttons to switch function.
- The PR 4510 is approved and certified as an add-on component for the PR 3000, 4000, and 9000 series of devices. All technical characteristics are valid with the PR 4510 attached.
Environmental Conditions
Operating temperature -20°C to +60°C Storage temperature -20°C to +85°C Relative humidity < 95% RH (non-cond.) Protection degree IP20 Installation in Pollution degree 2 & meas. / overvoltage cat. II Mechanical specifications
Dimensions (HxWxD) 73.2 x 23.3 x 26.5 mm Dimensions (HxWxD) w/ 4000/9000 unit 109 x 23.5 x 131 mm Weight approx. 20 g Common specifications
Supply
Supply voltage 6.5…20 V supplied from host 4000 / 9000 device Max. required power 0.15 W I.S. / Ex marking
ATEX II 3 G Ex ec IIC T5 IECEx Ex ec IIC T5 Gc FM, US Cl I, Div 2, Gp A, B, C, D T5; Cl I, Zn 2, Grp IIC T5 Cl I, Div 2, Gp A, B, C, D T5; Cl I, Zn 2, AEx ec IIC T5 Gc
FM, CA Cl I, Div 2, Gp A, B, C, D T5 Cl I, Div 2, Gp A, B, C, D T5; Ex ec IIC T5 Gc
Observed authority requirements
EMC 2014/30/EU & UK SI 2016/1091 ATEX 2014/34/EU & UK SI 2016/1107 RoHS 2011/65/EU & UK SI 2012/3032 Approvals
ATEX DEKRA 13ATEX0098 X IECEx DEK 13.0026 X UKEX DEKRA 21UKEX0167X c FM us FM22US0014X / FM22CA0009X CCC 2025322310006444 c UL us, UL 61010-1 E314307 DNV Marine TAA00000JD